This section contain instruments logs for the duration of the campaign. You can group the logs either by instrument or by date.

CIMEL sunphotometer

3rd April – CIMEL sunphotometer installation

CIMEL #181 sunphotometer is installed and registered as "FINO" AERONET station

5th April – CIMEL lunar/sun-photometer installation

CIMEL #989 (lunar/sun-photometer) is installed and registered as "FINO" AERONET station

CIMEL #181 still measures, but it is not registered in AERONET

8th April – Spider

A spider is removed at ~12:30 UTC

10th April – Spider behind the optical head

At ~7 UTC a spider was found behind the optical head of #181 (not inside the collimator unit) and was removed.

11th April – clouds

Almost overcast conditions

12th April – clouds

Overcast conditions

13th April – Clouds

Overcast skies today

17th April – Inspection for spiders

We found a spider web on #989, but not any spider

20th April – Clouds

Many clouds, almost throught the whole day

21st April – Clouds

Overcast skies today

22nd April – Clouds and rain

Clouds and rain today

29th April – Clouds and spider

7 UTC cirrus clouds in the sky

 

7 UTC a spider web was removed from #989 (but no spider was found)

PollyXT lidar

18th March – PollyXT arrival and installation

1200: RE, MK. PollyXT container is now installed in Finokalia station. Notes:
i) One A/C in operation due to malfunction of the vacuum pump.
ii) Water cooler is placed inside the container.
iii) Flashlamp counts set to zero.
iv) Air bubbles in the system cleared.
v) Pre-lasing test performed. Infrared light visible on the special paper at 1700 Volts.
vi) Laser beam alignment.
vii) Internet connection with the existing cellular network of Finokalia.
viii) 1064 off until it reaches the correct temperature.

19th March – 1064 channel activated

0845: mk. Restart polly.exe.
1232: mk. Plugged the 1064 cable. Restart polly.exe.
1242: mk. Beam down.
1318: mk. Internet is now provided by prepay vodafone (+306951124583). Compatibility with router and switch operation ould not be achieved. LTE modem directly plugged into PC.

Scanning polarization lidar

2nd April – Vertical measurements started

First vertical measurements were started during the evening. Measurements stopped due to a power failure.

3rd April – Vertical measurements continued

First set of vertical measurements collected. Comparissons with the Polly XT shows good agreement.

4th April – Vertical measurements

QA tests performed for alignment, triggerdelay and depolarization

Test measurements performed during evening. The data looks good.

 

5th April – Vertical measurements

Vertical measurements performed. Low clowds.

9th April – Scanning

First scans performed for three angles. 5, 30 and 60 degrees.

Multiple tests from 0 to 5 degrees performed.

10th April – Scanning

Measurements performed. Complex scanning scheme performed at 3 (three) angles (5, 30, 60 degrees) for N, S, E, W

13th April – Scanning measurements - preparation

The frame and motor for the scanning were installed. After a first set of vertical measurements, the lidar was installed on the scanning module.

RPG HATPRO Microwave Radiometer

1st April – Radiometer installation

radiometer installation finalized.

4th April – Instrument issues

The instrument stopped working during the night. Data aquisition restarted at noon

5th April – Instrument issues

The instrument stopped working during the night. Data aquisition restarted at noon. Measurement definition file changed to solve the instrument freezing issue.

12th April – Power related issues

the instrument stopped recording due to a power failure at the station. Measurements restarted.

PSR sunphotometer

4th April – installation

installed. it is cloudy so no aligment could be checked so it wasn't turned on. 

PR

5th April – alignment

8.32 UTC The instrument was aligned 4q 

the new offsets are

horizontal -0.047

vertical -0.092

 

11.00 UTC the instrument's interface was stucked for a couple of hours. it was restarted and functioned well

 

13 UTC optical parts cleaned

 

 

 

6th April – cleaning

12:00 utc

optical parts cleaned

 

software was stopped for an hour and a half for installing on the laptop pyranometer software

 

 

8th April – Optical parts cleaning

at ~12:20 UTC

9th April – Power supply

8:45utc

tracker loose power supply and stuck in a position towards the east.

the problem was solved and the tracker is pointing correctly again

global measurements weren't saved till today. problem solved

 

10th April – Optical parts cleaning

7 UTC: the optical parts are cleaned

14th April – cleaning

6-7 UTC optical parts cleaning

17th April – cleaning

9 UTC optical parts cleaning

also, until 9 UTC the instrument did not work, because a cable was detached

18th April – Optical parts cleaning

15:00 UTC optical parts cleaning

19th April – optical parts cleaning

16:00 UTC optical parts cleaning

20th April – Optical parts cleaning

17:00 UTC optical parts cleaning

24th April – Optical parts cleaning

9:40 UTC optical parts cleaning

27th April – Optical parts cleaning

10 UTC optical parts cleaning

CMP - UVB - Pyranometers

5th April – installation

13 utc 

installing both pyranometers is finished. some electronic parts are missing for correct connection to the logger. so tommorow the logger will be set up

6th April – soft installation

13 UTC the instruments are up and running

 

12 utc cleaning domes

 also today the logger software was installed

8th April – Optical parts cleaning

At ~12:20 UTC

10th April – Optical parts cleaning

7 UTC: optical parts cleaning

18th April – Optical parts cleaning

15:00 UTC optical parts cleaning

19th April – optical parts cleaning

16:00 UTC optical parts cleaning

20th April – Optical parts cleaning

17:00 UTC optical parts cleaning

24th April – Optical parts cleaning

9:40 optical parts cleaning

27th April – Optical parts cleaning

10 UTC optical parts cleaning

PREDE POM-01 sun-sky radiometer

4th April – Installation

Instrument installed

5th April – Prede POM allignement

The Prede POM was alligned at about 6:50 UTC

8th April – Spiders inside PREDE POM

Spider was removed from the tube of the instrument, at about 2:00 UTC

10th April – Optical parts cleaning

7 UTC: optical parts cleaning

14th April – cleaning malfunction

6-7 UTC

optical parts cleaning

during the optical parts cleaning and spider inspection a plastic ring from incide the collimator tube fell from the roof. We found it and we will put it back in place during a cloudy day

17th April – cleaning

9 UTC optical parts cleaning

18th April – Optical parts cleaning

15:00 UTC optical parts cleaning

19th April – optical parts cleaning

16:00 UTC optical parts cleaning

20th April – Optical parts cleaning

17:00 UTC optical parts cleaning

24th April – Optical parts cleaning

9:40 UTC optical parts cleaning